CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 12.1 Blood In-Line Sensor Check/Adjustment
3 - 153
3
11.If either check fails:
a. Loosen or Remove (if necessary) Blood In-Line Sensor Board screws to gain access to
Potentiometer R1.
Figure 3-82 Adjusting R1
b. With sample in the In-Line Sensor, adjust R1 until TP4 (+) 2.5VDC.
c. With blood in the In-Line Sensor, TP1 (+) must equal +0.3 to +3.25VDC. If not, replace the
Blood In-Line Sensor Board.
--- END ---