CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 10.2 Tube Positioning Arm Alignment
3 - 137
3
6. Ensure tube positioning arm(s) does not drag on top of platen. If necessary, loosen solenoid
bracket screws and adjust.
Only [OLD] F Loader with Clamp - (Others not Adjustable)
a. To raise arms slightly, loosen screws on solenoid bracket (
b. Push bracket up to give clearance.