CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 105
3
E Loader
[NON-CE MARK]
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Place Rack Guide Alignment Tool in right end of platen (
).
3. Load Mixing Head Alignment Tool in position #4 of rack.
Tool must be fully seated.