VERIFICATION PROCEDURES VP - 3.1 Platen Pre Check
3 - 18
3
VP - 3.1 Platen Pre Check
Purpose: Perform this procedure to ensure all hardware is secure, keyboard settings are correct,
and needle is home before performing Platen Checks / Adjustments.
1. Remove covers.
2. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
3. Set BAR (0/1) Switch to 1.
4. Set PRE rotary switch (if used) to 0 (painted dot faces 0).
5. Press (
↑
N) to ensure Needle is Home (X and Y axis will not move when Needle is not home).
6. Tighten all Platen, Pusher Arm, Sensor, and Laser Shield hardware (
7. Ensure Tower is secure and straight (side to side and front to back). If Tower cannot be
straightened, call Tech Support.
8. Go to
“VP - 3.2 Platen Wall Check” (Page 3-20)
Special Tools
DESCRIPTION
QTY
PN