CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 9.1 Sample Wash Block Check/Adjustment
3 - 125
3
VP - 9.1 Sample Wash Block Check/Adjustment
Purpose: Perform this procedure to verify and adjust Aspiration Needle Alignment.
1. Fully seat needle to collar and tighten clamp (
).
2. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
3. Set BAR (1/0) Switch to 1.
4. Set up check according to
.
5. Verify needle is centered in tool.
Special Tools
DESCRIPTION
QTY
PN
Needle Alignment
Tool
1
9420081
Rack Guide
Alignment Tool
1
9420084