28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 8.1 Needle Sensors Check/Adjustment
3 - 124
3
Figure 3-66 Sample Needle Home and Limit Clearances
9. Press (N
↓
) key to move Needle to the maximum DOWN position.
10.Verify distance between needle tip and Platen floor (
11.If check fails, adjust sensors.
--- END ---
F and G Home =
Flush or Recessed
to .040 in.
F
Limit = 0.5" (12.7mm)
G Limit=
1.0"-1.1"
(25.4mm-27.9mm)
Floor of Platen
SL-109B
F and G
LOADER