VERIFICATION PROCEDURESVP - 15.1 Barcode Reader Configuration (Europe Only) 3 - 163
3
VP - 15.1 Barcode Reader Configuration (Europe Only)
Purpose: Perform this procedure to verify and edit Barcode Reader configuration, Europe Only.
CAUTION:This procedure supersedes ACS product info #97-005, CD 012A and 97-H-1996-007.
1. Prepare Link Cable (use
as a reference).
NOTE:Cable compatible with [NON-CE MARK] and [CE MARK] Barcode Readers.
a. Use four-wire signal cable about 2m long.
b. Attach connectors (DB9 - female) to each end.
c. Label connectors BCR and COMPUTER.
2. Turn power OFF at Sample Loader and Data Station.
3. Remove Sample Loader cover(s) to access CPU Main Board.
4. Disconnect Barcode Reader Cable at CH2 on CPU Main Board.
5. Connect BCR connector of Link Cable to Barcode Reader Cable.
6. Connect COMPUTER connector of Link Cable to port labeled EXTERNAL COMPUTER (back of
Data Station) or COM1 of laptop.
Special Tools
DESCRIPTION
QTY
PN
[US] N/A
[Europe]
VBCDIAG Program
1
N/A