28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.5 Rack Clearance Adjustment
3 - 28
3
VP - 3.5 Rack Clearance Adjustment
Purpose: Perform this procedure to adjust Rack to Beam Block clearance,
1. Perform Check procedure
“VP - 3.4 Rack Clearance Check” (Page 3-26)
.
2. Loosen (2) screws (
) inserted through bottom of Platen below Beam Block. If screws
cannot be accessed with Platen installed, remove Platen (
3. See
and
for proper rack clearance.
NOTE:DO NOT loosen Rack Guide(s).
4. Adjust Beam Block and tighten hardware.
NOTE:If Platen is removed, DO NOT install at this time.
5. Perform Check procedure
“VP - 5.1 Tube Detect Sensor (Beam Block) Check” (Page 3-66).
6. Go to
“VP - 3.6 X-Axis Solenoid to Guide Rod Check” (Page 3-31)
.
Special Tools
DESCRIPTION
QTY
PN
Rack Guide
Alignment Tool
1
9420084
Feeler Gauge Set
1
1-43627-01