3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.17 X-Axis Solenoid to Rack Guide Adjustment
3 - 58
3
f.
Check Left Rack Guide in same manner. Remember to stop Solenoids before driving to left
(Limit) Sensor.
g. Repeat check (
“VP - 3.8 Rack to Laser Shield Check” (Page 3-39)
.
h. Repeat Steps 2a-2g for Left (Limit) Sensor.
Figure 3-26 X-Axis Sensors
--- END ---
Right End
Sensor
(Home)
Left End
Sensor
(Limit)
Slowdown
Sensor
DO NOT
ADJUST
Loosen (2)
Allen Screws
(2.5mm)
SL-173B