DICOM Configuration Parameters:
Definition:
Log Level
Defines the level of detail information that will be displayed in the DICOM
Queue.
Maximum entries cached
Enter a number between 0 and 999
Set according to department requirements. A higher number provides access to
more scheduled exams, but increases the time required to search for the exam of
interest.
Minimum/Maximum Film Density
For defining Minimum/Maximum density of images on film. Value entered is
expressed in hundredths of OD (optical density). A value entered of 150,
corresponds with 1.5 OD. Typical acceptable values to be entered are 000-399,
with 020 (for “Min”) and 300 (for “Max”) being a potential starting point.
Network Type
Allows for manual or auto setup of network speed and duplex operational state.
Selections include: Auto-Negotiate, 100MB/Full-Duplex, 100MB/Half-Duplex,
10MB/Full-Duplex, 10MB/Half-Duplex
Nickname
e.g., “Prep Room Printer”
An informal name to identify the device to users in the department. This name is
typically assigned in the department.
Network Usage Type
Defines the interval at which DICOM images will be transferred over the network
to the PACS storage device. Either immediately upon each press of STORE
(“Exam in Progress”), or at the termination of the current examination (“Exam
Completed”).
-
Exam In Progress
-
Exam Completed
No Color
If selected (checked), images that are captured when the
Z.ONEULTRA SP
is
being operated in modalities where no “Color” is present (B-Mode, PW Doppler,
M-Mode), they will be transferred to the receiving DICOM device (Network
Store system) as “
Grayscale-Monochrome2.
Despite this “
No Color
” selection being active, all images captured while in Color
Doppler mode will still be transferred to PACS using the color map specified in
the
Color Model
field.
If unchecked, all images (regardless of operating mode) will be sent to PACS
using the color map specified in the
Color Model
field.
PDU Receive/Send Size
The size (in bytes) of the basic data packets (PDU: Process Data Unit) that will be
trasferred during receive and transmit operations. (default is 16384 = 16K)
Pixel Spacing
Adds ability to assign CT/MR-specific pixel spacing to ultrasound images sent to
PACS, by assigning non-US pixel spacing from the 2D US region
Polling Interval (minutes)
−
None
−
choose minutes
The frequency of automatic updates to the worklist. A short polling interval may
impact system performance. Choose a preferred interval; or choose
None
to
update the worklist only on demand.
Port number
The TCP/IP port number that is being used by the DICOM device connected to
the
Z.ONEULTRA SP
. Default port value is 104.
Queue Timeout (sec)
The maximum amount of time that a DICOM job will remain active in the queue,
before it times out.
Queue Start Type
Defines whether the start of jobs in the DICOM queue will be performed
automatically
or started
manually
.
Reconciliation
For exams
NOT YET
sent to PACS. Specifies that the system will interrogate the
patient information, prior to sending that exam to DICOM PACS device. IF any
images in that exam have differing information, the system will reconcile
(correct) those values prior to sending.