Q00180 Rev B
Page 148 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
16.7.5 MANUAL Log File Capture (Store) – Download to ZONARE
1.
Using the trackball, position the arrow cursor over the
Store Current Logs
box
2.
Press the “SET”
button to initiate the creation of the LOG directory and the writing
(“Capture”) of the log files to the internal CompactFlash storage media. The directory and
files will look similar to that shown below:
3.
Using the trackball, position the arrow cursor over the
Send System Logs
box, and Press the
“SET”
button to initiate the system log transfer
16.7.6 BRUTE-FORCE Log File Capture – (System Problem Condition)
1.
Holding down (extended press for approx 5+ seconds) the “
SERVICE
” key, on the upper
row of the QWERTY keyboard on the SmartCart SP, will result in manually generating the
capture of a set of internal “LOG” files,
16.7.7 MANUAL Image File Capture (Store) – Download to ZONARE
NOTE
:
The following process is used for capturing (and storing) an ultrasound image screen
snapshot as a .BMP file to the”
LOG
” directory of the internal storage inside the Scan
Engine. The intent is for capturing, and then sending (via an FTP transfer) digital images to
Zonare for evaluation
Operate the system normally until the desired ultrasound image event has occurred, at which
time press “Freeze” to retain the image on the screen
.
Содержание Z.one Ultra SP