CAUTIONS
During immersion disinfection, never immerse the transducers longer than the maximum
specified time (as per the Operator Manual instructions) of 45 minutes. Damage may occur to
the transducer housing and/or components if disinfection times exceed these recommended
limits.
Using a non-recommended cleaning or disinfectant solution, incorrect solution strength, or
immersing the transducer deeper or longer than indicated can damage the transducer.
To prevent possible damage to the electronics of the transducer, never immerse the transducer
beyond the point (10mm below the top edge of the transducer housing), as shown in the figure
below.
Figure 46: Transducer immersion limits
1.
Disconnect the transducer from the
Z.ONE
Ultra
.
2.
Remove the transducer cover/sheath and discard.
3.
Clean the surface of the transducer, using a soft cloth dampened with mild soap (or isopropyl
alcohol) to remove particulate, ultrasound coupling gel, or body fluids. Apply the cleaning
solution to the cloth, and then use to wipe down transducer.
4.
After cleaning, wipe the transducer again with a cloth soaked with clean water to remove
solution.
5.
Mix the disinfectant solution, following the disinfectant label instructions for solution strengths
and disinfectant contact duration (Refer to Error! Reference source not found.)
6.
Immerse the transducer into the disinfecting solution, ensuring it is not immersed below the
depth indicated in the figure (Refer to
Figure 46: Transducer immersion limits
)
10mmm
10mmm