12.
Remove the USB Memory Stick, and press the “SELECT”
key again to finalize the software
installation. The system should perform an automatic reboot, powering back up in a fully
operational condition
16.5 SCAN MODULE: “Clean” Software Install Procedure (“Undocked”)
ATTENTION
The Scan Module requires connection to an (optional) AC Power Adapter in
order to be operated for software installation while in an “undocked” (outside of
the Cart) condition.
All software installations performed while in a “standalone” mode, will be
automatically be executed as a “Clean” install (no user settings backed-
up/restored). All user Presets and System settings MUST have been backed-up to
USB Memory Stick, PRIOR to performing this form of software installation on the
Scan Module.
1.
Attach a Zonare AC Power Adapter (option) to the DC power input port on the rear of the Scan
Module, or have the Scan Module physically docked on Scanner Deck of SmartCart.
2.
IF Scan Module is being powered by docking it in the SmartCart, make sure that AC is power
present at rear of Cart, and the circuit breaker in the “
1
” / ON position (or Z-PAK battery pack is
installed in system)
NOTE
: The Cart will NOT be actively “turned on” (power button on Scanner Deck will
NOT
be pushed) during this direct Scan Module “Clean” software installation.
3.
Insert the USB Memory Stick (containing system software installer files) into the USB port at the
rear of the Scan Module, pressing lightly to ensure it is seated in the connector
NOTE
: IF the USB Memory Stick type being used is too long to fit in the clearance between
the back of the Scan Module and the gray plastic rear panel on the Cart, place a flat-
blade screwdriver under the sliding panel, to raise it up a sufficient distance to allow
clearance.
4.
Press the Power ON/OFF button on the Scan Module, and allow it to power up.