Q00180 Rev B
Page 72 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
Selecting the “
REBUILD
” box will regenerate the database’s index file (directory) and
potentially resolve the problem. Select the desired media device, using the provided pull-down
menu, prior to initiating the “Rebuild” process.
--------------------------------------------------------------------------------------------------------------------------
•
Erase All Patient Data:
WARNING
Initiating the “
ERASE
” function will completely delete ALL patient exams (all stored images)
on the internal ARCHIVE storage of the Scan Engine.
DO NOT
perform this function unless
the system’s User has confirmed that all needed exams/images have been exported/transferred
to another medium, or are no longer needed.
Due to this operation performing individual “deletes”, the “ERASE” operation may take an
extended period of time to complete if the selected Archive media contains an extensive number
of patient exams.
In extreme cases where the system is totally unable to access the ARCHIVE on either the scan
engine or other media (HDD in Cart, etc), selecting the “
ERASE
” function will completely
clear (erase) the following:
•
All stored patient exams (images)
•
All patient exams pending DICOM transfer in DICOM QUEUE (if C drive selected)
Select the desire media device using the provided pull-down menu.
--------------------------------------------------------------------------------------------------------------------------
•
Format Storage Media:
NOTICE
Initiating the “
FORMAT
” function will result in removing ALL of the target media device.
This includes all of the following IF the “C” drive is selected:
•
Patient exams
•
System Logs
•
DICOM Queue jobs
•
Some Transducer file
To prevent Users from accidentally performing this operation a “Service Password” prompt will
appear prior to this “FORMAT” operation activating. Contact Zonare Technical Support in order to
obtain the service password and assistance on this operation.
Due to the erasing of some transducer files, reloading of system software will be required after
performing this “FORMAT” operation.
In extreme cases where the system is totally unable to access the internal storage media of the
scan engine (“C” drive) or other media (HDD in Cart, etc), selecting the “
FORMAT
” function
Содержание Z.one Ultra SP