Q00180 Rev B
Page 52 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
NOTE
:
The SmartCart SP has the ability to be powered by AC or battery (IF the Z-PAK
battery pack “Option” is ordered). When the SmartCart SP is fully charged the
battery should allow for normal system operation for more than 1 hour.
When the scan engine is docked into the SmartCart SP, its internal battery is
subject to being charged. If the SmartCart SP is being AC powered the Scan
Engine will utilize the main power supply, with source power coming from the AC
input. If the SmartCart SP is running on the Z-PAK battery the scan engine will be
charged using source power derived from the battery in the SmartCart SP.
2.
Locate the power button on the front/left of the scan engine Scanner Deck. Press and release
the On/Off button to power on the
Z.ONEULTRA SP
.
3.
The
Z.ONEULTRA SP
will take less than a minute to complete the normal power on sequence.
5.6.5 Powering-OFF
the
Z.ONEULTRA SP
System
CAUTION
To ensure the integrity of the stored User parameters, and system configuration, the Power
ON/OFF button on the Scan Engine should always be used
FIRST
, as the method for shutting
down a fully operating
Z.ONEULTRA SP
(SmartCart SP and Scan Engine) system.
Failure to allow for a normal power-down sequence by the Scan Engine (such as by improperly
pulling out the AC power cord or toggling the AC circuit breaker on the SmartCart SP while
the
Z.ONEULTRA SP
is still operating), may result in lost patient image data.
5.6.5.1
Scan Engine (standalone) Power-OFF
STEPS:
Содержание Z.one Ultra SP