Q00180 Rev B
Page 136 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
Description of Key Functions
Key
Function
Operation Performed
Back
Continue
Pressing this key results in the Z.ONEULTRA SP exiting the user break menu, and
performing a normal system boot operation.
Set
Safe Mode Boot
Not implemented at this time
Select
Install Software
Installs software from USB Memory Stick, retaining all existing User Presets and System
(configuration) Presets (by automatically backing up to USB Memory Stick, prior to
installing new software).
Tab
Export system logs
Copies content of internally stored system LOG files, to externally inserted USB memory
stick
Backup System Data
Moves the contents of the internal exam archive storage (Patient Exams) to the currently
attached USB Memory Stick.
Description of Optional Hidden Functions –
(4.x Software)
Key
Function
Operation Performed
Access Key(s)
Clean SW Install
Installs software from USB Memory Stick, over-writing
(erasing) existing User Presets and System (configuration)
Presets, with Zonare default values. NO formatting of the media
is performed.
Optimize
Factory Clean Install
The “C” and “D” partitions of the internal media of the Scan
Engine are completely reformatted (all content erased), as a
precursor to installing system software from USB Memory
Stick. As in the normal Clean Install, this process erases all
existing User Presets and System (configuration) Presets;
replacing them with Zonare default values..
Optimize -> Measure
->
Erase User Data
Completely erases
ALL
stored patient exams (Image
ARCHIVE). Additionally erases all entries in the DICOM
Queue and LOG file directory, on the internal CompactFlash
media inside the Scan Engine.
Optimize -> Measure
->
Содержание Z.one Ultra SP
Страница 1: ......