System Verification
MTP Bolster Plate Access/Removal-Replacement:
STEPS:
1.
Remove the two (2) Phillips head screws securing the right side of the lower cosmetic cover.
The screws are located on the right side next to the printer mounting bay
2.
With the right side lower cosmetic cover loosened, grasp the gray plastic Right CEE cosmetic
panel (that surrounds the mock transducer storage ports) and remove it
3.
Remove the three (3) Phillips screws securing the MTP Bolster Plate to the cart
4.
Disconnect the Power cable at the rear of the MTP Bolster Plate.
5.
Disconnect the MTP dongle cable from the rear of the MTP Bolster Plate.