Q00180 Rev B
Page 57 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
11.
During the “Recalibrate” process, the left status LED flashes ORANGE. Once the
“Recalibrate” process is completed, this same LED will intermittently flash between ORANGE
and GREEN.
NOTE
: - The “Recalibrate” process may require 14 hours to fully complete
- The battery pack should
NOT
be removed
prior
to completion of the “Recalibrate” process,
as the battery will be at an unknown level of charge
5.6.7.2
Charging Scan Engine Battery Packs
STEPS:
12.
Connect the AC power cable to an active wall outlet, and connect the DC power cable between
the power supply and the 2-Bay Battery Charger
13.
Insert the battery pack into the EITHER bay of the 2-Bay Battery Charger
14.
The charging cycle on the docked battery pack will automatically begin to execute
15.
During the “Charging” process, the corresponding status LED flashes GREEN. Once the
“Charging” process is completed, this same LED becomes solid GREEN.
NOTE
: -
The
“
Charging
” process, in the two bays of the unit, is performed in a “sequential” fashion. If
two battery packs are docked into the 2-Bay Battery Charger at the same time, the charger
will perform the complete charging sequence on one battery pack, before starting the
charging sequence on the second battery
.
Status LED Condition
LED
Definition
OFF
No battery pack detected in bay
ORANGE – “Flashing”
“Recalibration” of battery pack in progress
ORANGE/GREEN –
“Flashing”
“Recalibration” complete (fully charged)
GREEN – “Flashing”
“Charging” of battery pack in progress
GREEN - Solid
“Charging” of battery pack complete
ORANGE - Solid
“Standby” – Battery in bay, waiting to be charged
RED – “Flashing” (Error)
Error Cause
Resolution
- Foreign object in bays
Clear charging area
- Battery pack defective
Replace battery pack
- Excessive battery temp
Remove battery to cool
[65
o
C / 149
o
F]