Q00180 Rev B
Page 203 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
•
#2 Phillips Screwdriver
Procedure
Battery Pack Removal:
STEPS:
1.
Perform removal process as described in Z-PAK Battery Plate removal procedure (included
previously in this section of the manual).
Power Supply Module Removal:
STEPS:
1.
Ensure the
Z.ONEULTRA SP
is powered “OFF”.
2.
Place AC circuit breaker, located at the rear of the SmartCart SP, in the “
0
” (OFF) position.
3.
Disconnect the main AC power cord from the rear of the SmartCart SP or unplug from the wall
source.
4.
Undock the scan engine from the Cart, and set aside.
5.
Put the User Interface in the highest position.
6.
Set the front caster brakes to prevent system from rolling.
7.
To allow for removal of the AC power cable (used to power an on-board USB printer device,
and an item that is included with the complete Power Supply Module) from the SmartCart SP,
the lower access cover panel will need to be removed.
8.
Remove the two Phillips-head screws that retain the lower access cover panel, in the printer
mounting bay area. This panel is located under the scanner Scanner Deck, at the front of the
SmartCart SP.
9.
Cut any plastic tie-wraps that retain the AC power cable to the chassis, and free it from its
mounting for later removal with the Power Supply Module.
10.
The Power Supply Module is located under the system. For the easiest access, tilt the system
so it rests on the front handles, exposing the bottom of the unit.
11.
Remove the eight (8) Phillips-head screws retaining the power supply module to the bottom of
the SmartCart SP and remove module.
Содержание Z.one Ultra SP