Q00180 Rev B
Page 160 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
NOTE 2:
If the cable is
too tight
(the lift is not staying locked in position when the release
lever handle is fully released – aka “drifting”), rotate the adjuster barrel in the
clockwise
(opposite to what is shown in the illustration above) direction just far
enough to resolve the “drifting” condition.
2.
Reverse cable access process above, to complete this procedure
18.6
Touchscreen Calibration - (Scan Engine)
IMPORTANT
The touchscreen is very sensitive to the “technique” used by the Operator in making
contact at each of the calibration points with the stylus. It is very important that only a
single discrete (and firm) contact be made at each of the green splash graphics
calibration points. Any excessive debounce, when pressing the stylus to the screen, will
likely result in a failure of the calibration process.
18.6.1 Calibration Procedure
1.
Remove the Scan Engine from the SmartCart SP
2.
Press the On/Off switch on the Scan Engine, to power-on the
Z.ONEULTRA SP
3.
Using the “
Tab
” button and
Menu Control
on the Scan Engine user interface, advance to the
“
Tools
” menu
and then sequence through the following selection:
Tools
tab
--> SYSTEM SETUP --> Touchscreen
4.
Press the “SELECT”
button on the user interface. The Touchscreen calibration screen
should appear, with the first (of 3) green “splash” target points being should now be displayed
1st
2nd
3rd