Q00180 Rev B
Page 163 of 244
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
6.
Visually inspect the output of the LCD Display, to confirm the desired video levels are
achieved. Repeat the above steps, if required
Grayscale Test Pattern:
1.
Using the “
Tab
” button and
Menu Control
on the user interface, advance to the “
Tools
” menu
2.
Sequence through the following selections to bring a grayscale test pattern (called “
Bar
Pattern
” in the menu) onto the SmartCart SP display.
TOOLS
tab
--> SYSTEM SETUP --> DISPLAY --> BAR PATTERN
.
3.
To remove the bar pattern from the display; push the “
BACK
” button
on the user interface.