OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Macro Menu
LSM 510 META
DuoScan
4-186
B 45-0021 e
03/06
4.8.4.5 Excitation
Fingerprinting
System requirements
:
−
LSM 510 NLO or LSM 510 Meta NLO system in combination with a software tunable ultra fast
laser, software release 3.2 or higher.
With the LSM 510 NLO and LSM 510 META NLO in combination with a software tunable ultra fast laser
(multiphoton lasers such as Chameleon from Coherent or Mai Tai from Spectra Physics) one can acquire
excitation spectra and use these spectra to perform linear unmixing. This process is called Excitation
Fingerprinting.
To use the functionality, a macro is provided to perform the necessary steps for the calibration of the
system and the image acquisition.
The calibration of the system ensures a constant laser power at the specimen plane by adjusting
the AOM transmission in coordination with the applied wavelength. This has to be done for
each set up in the field where the laser is also available.
(1) Set up the following configuration for
calibration and image acquisition:
•
Use the KP 650 as main dichroic and the
KP 685 as emission filter.
•
Select excitation laser line and set transmission
values mode.
•
To perform the AOM calibration put the meter
in the specimen plane using the objective and
main dichroic you will use for image acquisition
later on.
•
Select Frame as the scanning mode.
•
Use the bidirectional scanning mode.
•
Adjust the zoom to the maximal value (40) by
moving the slider to the far right position.
•
Set the scan speed to Max
•
Start the Macro
•
Start scanning with
Scan Cont
.
•
Change view to
AOM calibration
.
•
Chose lowest wavelength available.
•
Set AOM Transmission to a value around 8.
•
Activate the
Auto+
option checkbox (Auto-is chosen when starting with the highest wavelength
available).
•
By pressing
Add and + 10 steps
the AOM Value will be added to the list and the laser is tuned to the
next wavelength: 10 nm further.
Fig.
4-180
Excitation Fingerprinting macro
window
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......