OPERATION IN EXPERT MODE
LSM 510 META
Maintain Menu
Carl Zeiss
03/06
B 45-0021 e
4-233
Non-regularities of the scanner feedback.
The ticks on the outer sides of the grid vary about more than 1 tick width between consecutive
images (in the middle of the calibration process, the linearity is optimized and the problem starts
to occur).
•
Stop the calibration procedure.
•
Call the LSM service hotline.
If optical calibration comes not to a successful end, please contact your service hotline.
4.10.1.4
Scanner Calibration in LSM 5
Software
In LSM 5 Software the
do full calibration
(Measure spectral response)
checkbox must be
set at the first calibration procedure
Speed 11-13
(optical)
.
Fig. 4-229
Scanner Calibration window