OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-106
B 45-0021 e
03/06
(1)
Start via Trigger
For the start via trigger control (
Trigger
button
activated), first determine the trigger key which is
to trigger the start of the Time Series.
•
Open the
Trigger in
list box with a click on the
arrow button.
•
Choose one of the trigger keys 1 to 4 (e.g.
Trigger1
).
It is also possible to trigger an out signal via trigger control.
•
Open the
Trigger out
list box with a click on the arrow button.
•
Choose one of the trigger keys 1 to 4 (e.g.
Trigger1
).
In this example, the scan procedure is triggered on pressing key
1
of the trigger control, and an out signal
is given at the same time.
When starting a Time Series via Trigger, the
Start T
or
Start B
button must be pressed first.
Waiting for Trigger
will then be displayed in the status line.
Then the relevant trigger key on the Trigger Control must be pressed to start the first scan
procedure of the Time Series.
(2)
Start via Time
For the start via the time set on the PC (
Time
button activated), the start time must be entered
first in the
Time
input box.
•
Click in the
Time
input box to open it.
•
Enter a start time via the keyboard. Then click
outside the input box once to close it again.
When starting a Time Series via the time, the
Start T
or
Start B
button must also be pressed in
this case.
Waiting for Start Time
will be displayed in the status line.
The Time Series is started when the starting time has been reached.
The starting time for the Time Series can be changed online.
(3) Pre-Scan
The
Pre-Scan
allows to image the sample continuously before actually starting the
image acquisition.
This function should only be used for starting the time series manually.
After clicking the Start or Start B button, the button
Go
appears on the
lower right hand side of the control window. By clicking this button the
actual image acquisition is started.
Fig. 4-85
Start Series panel
Fig. 4-86
Start Series panel
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......