In case the system is not operating following the removal or attachment of any device on a port
with safety interlock check again the connection of the Contact ring to the Sensor ring.
For dismounting the lamps, TPMT or switching
mirror slightly unscrew the contact ring first
2
2
) so it can be moved
away from the sensor ring (Fig. 1-11/
1
). Then
unscrew the lamp, TPMT or switching mirror (main
screw
Fig. 1-11/
) which is in one of the recesses of the
2
). Hold the device to be
dismounted with one hand while unscrewing to
keep it from dropping. The now empty port has to
be closed with the blind cap to restore the
functionality of the system.
Use the main screw of
the port to fix the cap. Make sure the pins of the
cap depress the sensors of the sensor ring.
Do not remove the sensor ring from
the microscope. This might result in
failure of laser safety and a non
operating system.
For mounting any lamp, TPMT or the switching
mirror back onto the microscope reverse the steps
for dismounting the device. Be careful not to bend
the pins on the contact ring when screwing the
device onto the microscope port.
For the Axiovert 200 M transmission port and the
two ports available on the motorized switching
mirror no sensor ring is present. Instead the
sensors are directly installed at the Axiovert 200 M
transmission port or the two ports of the
motorized switching mirror.
1.11.2
Mounting and Dismounting the
Scan Heads
The scan heads are connected to the microscope
via an integrated safety interlock. They can be
moved between two microscopes. Make sure the
system is shut off completely before starting the
following procedure:
Be aware that the scan head weights up to 15 kg.
Fig. 1-11
HBO lamp mounted to
Axiovert 200 M with sensor
ring and contact ring for laser
safety
Fig. 1-12
Blind cap for closing any port
equipped with an interlock device
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......