OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-82
B 45-0021 e
03/06
Tabs
Z Sectioning
:
Tab for setting of
Number of Slices
,
Interval
and
Current Slice
via slider / arrow
button.
Mark First/Last
: Tab for determination of the Z-value for the first and last XY-image of the stack,
combined with manual focusing or
Stage control.
Hyperfine
Z Sectioning
:
Tab for production of a Z Stack using the optional Piezo focus for objectives.
First
:
Scanning / Display of the beginning (first XY-image) of the stack.
Mid
:
Scanning / Display of the center (XY-image in the center) of the stack.
Last
:
Scanning / Display of the end (last XY-image) of the stack.
Refr. Corr.
:
Considers the different refractive index between the immersion medium of the
objective (n') and the embedding medium of the specimen (n), which can be set
between 0.5 and 3 via the slider / arrow buttons
'
n
n
Ratio
=
X:Y:Z=1:1:1
Clicking on this button will set the Z-interval in such a way that the voxel has identical
dimensions in the X-,Y- and Z-directions (cube).
Auto Z Corr.
This function permits the set values of the scan parameters
Detector Gain
,
AOTF
,
Ampl. Offset
and
Ampl. Gain
(as measure for the brightness level) to be varied
between two freely selectable slices of a stack to be recorded. During the scan
procedure, the interim values of these three parameters are automatically linearly
interpolated between the initial and end values (see page 4-88).
Fig. 4-63
Optical Slice window
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......