LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-69
Control buttons for reference spectra
RS1 ... 8
Menu for selecting a display color and a reference spectra (reference spectra
derive from earlier experiments via mean of ROI or ACE tools); selected spectra
and colors appear in the wavelength chart. Unmixed results will be displayed
during scanning, lambda stack will neither be displayed nor stored.
META detector
slider
Definition of number of PMTs out of the 32 PMT Detectors to fit the required
spectral range. To optimize acquisition speed, a range detected in 1 or 2 passes
is recommended
.
Start
box
Start value of required spectral range.
End
box
End value of required spectral range.
Step
box
Step width in spectral range (10.70 recommended).
4.5.3.10
Non Descanned Panel
The functions of the
Non Descanned
panel are described in CHAPTER ANNEX of the printed
manual.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......