LSM 510
3D FOR LSM
LSM 510 META
Functions
Carl Zeiss
03/06
B 45-0021 e
7-33
Segment - Interactive
This function carries out a grey value segmentation by means of thresholding.
Fig. 7-24
The
Interactive
tab sheet of
Segment
dialog window must be selected.
Segmentation is especially used to generate binary regions. These are required for the measurement.
Two threshold values determine which grey value range of the
Input
image sequence is preserved and/or
deleted in the
Output
image sequence. Only one channel of a multichannel sequence can be selected as
Input
.
Output
will always be a single channel sequence.
The vertical scaling of the histogram can be adjusted with the scroll bar at the right edge of the
histogram. This setting has no influence on the function.
The thresholds
L
ow and
H
igh are determined either by moving the borderlines in the grey value
histogram or by the scroll bars underneath. Furthermore, the values for
L
ow,
C
enter and
H
igh can be set
through entry in the corresponding fields.
To move the lower (
L
) and upper (
H
) thresholds at the same time, move the vertical line in the grey value
histogram or set the scroll bar (
C
).
The
Green
and
Blue/Red
option buttons of the parameter
Colour
determine whether the voxels within
(
Green
) or outside (
Blue/Red
) of the grey value interval [
L, H
] are displayed with the corresponding
colour.
If
Green
is selected, the voxels within the selected interval are highlighted in green. The rest of the image
retains its original grey values. The voxels with the grey values
L
ow and
L
ow+1 are displayed in blue. The
voxels with the grey values
H
igh and
H
igh-1 are displayed in red.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......