LSM 510
NOTES ON DEVICE SAFETY
LSM 510 META
User Interface
Carl Zeiss
03/06
B 45-0021 e
1-25
To ensure functioning of the system and laser safety the following connections have to be changed:
1. The connection of the microscope to the safety interface of the system is located either on the
additional Safety-Box (Axioskop 2 FS
MOT
1
) or on the rear side of the microscope (Axiovert
200 M or Axio Imager.Z1; Fig. 1-23/
1
1
). This connection has to be unplugged from
the microscope which is not in use and plugged into the microscope to be used following the
exchange of the scan head.
Fig. 1-22
Safety-Box of Axioskop 2 FS
MOT
with main connection to safety
interface (1)
Fig. 1-23
Connection of Axio Imager.Z1
to safety interface (1) and
electronics (2)
1
Fig. 1-24
Connection of Axiovert 200 M to
safety interface (1)
Fig. 1-25
Four CAN connections (1) are
available on the rear side of the
electronics module. The microscope
in use has to be connected to one
of them
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......