LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Maintain Menu
Carl Zeiss
03/06
B 45-0021 e
4-227
(2) Calibration conditions
Before the calibration process can be started, the system has to be in operation for at least one hour.
(3) Calibration procedure
•
Click on the
Scanner
button in the
Maintain
subordinate toolbar of the
Main
menu.
−
This opens the
Scanner Calibration
window.
•
Click on the Speed 1-10 (electr.) or Speed 11-13 (electr.) button respectively.
•
For electrical calibration of speeds 11-13 the appropriate zoom factors have to be applied (11 : zoom
≥
2.5, 12 : zoom
≥
3.6, 13 : zoom
≥
5.6).
•
Activate the Display Graphics check box enables to check the progress of the calibration process on
the Progress Status bar.
−
During successful calibration process, the status button is of green color, in case of an error it
switches to red. The progress of the calibration process is indicated by the
Progress Status
bar.
The calibration process is completed, when the indicator button is grayed.
•
Click on the
Calibrate
button to start the automatic scanner calibration.
•
Confirm warning information with
OK
.
•
Click on the
Close
button to close the
Scanner calibration
window.
The
More
function is for servicing purposes only and can only be performed by authorized personnel. Its
access is therefore password-protected.
4.10.1.2
Optical Calibration with Speed 11-13 (Bidirectional)
(1) Preliminary
notes
The optical calibration of scan speeds 11-13 (bidirectional) can only be performed at systems with
complete hardware level of Release 2.8.
The optical calibration procedure has to be repeated every two weeks in normal use and after long delay
times of the system.
The minimum duration of the calibration process is 10 minutes. However, it can last up to a maximum
time of 40 minutes depending on the performance of the scanner in use and the actual tuning
conditions.
If the optical calibration is successfully finished, there is no need to start the electrical calibration for
unidirectional scanning of speed 11-13 (in opposite, the electrical calibration would overwrite the much
more accurate values of the optical calibration procedure).
With speed 11, 12 and 13, scanning is performed at scanner frequencies of 868, 1042 and 1306 Hz,
respectively.
In bidirectional scanning, the total line frequencies are 1736, 2084 and 2612 Hz, the image recording
times for 512 lines are 0.29, 0.25 and 0.20 seconds.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......