LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-105
Pause
button
Interrupts the Time Series. Button labeling is changed to
Resume
. A current scan
is performed until the end. When the button is pressed again, the Time Series is
immediately continued with the next scan procedure.
Bleach
button
Starts a Bleach procedure without a Time Series. Bleach procedure must be
defined first in the
Bleach Control
window.
Mean ROI
button
Creates a Time Series with the intensity values of the Frame or the default ROIs. An
average value is formed of the intensity values of the Frame or the ROIs
determined in the relevant scan procedure. These average values are displayed in
an extended
Image Display
window as a function of the time which has passed.
The status line, in which the phases of the current Time Series or notes for the user are displayed, is in the
lower part of the
Time Series Control
window.
4.5.6.2
Start Series Panel
In this panel, the parameters for the start of the
time series are set.
The following functions are available:
Manual
button
The time series is started manually with a click on the
Start T
or
Start B
button.
Trigger
button
The time series is started via a trigger signal from Trigger Control.
Time
button
The time series is started when the set time is reached. The internal computer time
applies.
Time
input box
Input of the time for the start of the time series (
Time
button activated).
Trigger in
list box
Selection of the trigger key (1-4) with which the start is to be triggered (
Trigger
button activated).
Trigger out
list box
Selection of the trigger keys (1-4) for the out signal.
Pre-Scan
If this box is checked before starting the Time Series a continuous scan is started
but no images are acquired until the button
Go
, which then appears on the
lower right hand side of the control window, is clicked.
Fig. 4-84
Start Series panel
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......