NLO Non Linear Optics Laser for LSM 510 NLO and LSM 510 META NLO
The LSM 510 can be operated with a Titanium-Sapphire-Laser to perform Multiphoton Microscopy.
Depending on the type of laser the following operation procedures apply:
•
The operation of software controlled lasers (Chameleon laser from Coherent or Mai Tai laser from
Spectra Physics) is integrated into the LSM software. Operation of Chameleon laser (Coherent; For
further details please refer to the operators manual of the laser)
−
Operation of Chameleon laser (Coherent; for further details please refer to the operators manual of
the laser)
The laser should be kept in Stand by mode for daily or weekly operation. The cooling of the laser
must not be shut down during stand by. Only in case of more than two weeks without operation
the laser should be shut down completely. This has to be performed according to the operation
manual for the laser provided by Coherent. The shut down takes about an hour as does the start
procedure afterwards. The key on the front panel of the laser power supply is used for switching
between Stand by mode and On mode.
The laser has to be in On mode to allow the LSM software to operate the laser.
−
Operation of Mai Tai laser (Spectra Physics; for further details please refer to the operators manual
of the laser)
The laser should be kept in Stand by mode for daily or weekly operation. The cooling of the laser
must not be shut down during stand by. Only in case of more than two weeks without operation
the laser should be shut down completely. This should be performed according to the operation
manual for the laser provided by Spectra Physics. The shut down and the start procedure take up to
one hour. The laser is put to On mode or Stand by mode using the LSM software. Therefore no
further operation of the laser via the laser power supply unit is needed. The key on the front panel
of the laser power supply is an interlock key, when removed the system cannot be operated. The
power switch should be kept on for leaving the laser in Standby.
•
Operation of Millenia Tsunami (Spectra Physics) or Verdi Mira lasers (Coherent)
These lasers are not operated via the LSM software. Refer to the operators manual of the
manufacturer for using these lasers in combination with the LSM 510. However it is needed to
indicate the tuned wavelength in the software to ensure the appropriate setting of the AOM
frequency (see below).
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......