LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Process Menu
Carl Zeiss
03/06
B 45-0021 e
4-151
4.6.12.4 Ratiometric
Dyes
•
Fura-2, Indo-, SNARF, Cameleon, Ratiometric Pericam, Phluorin,...
•
Display fluorescence ratio R over time
•
Display fluorescence ratio R corrected for background/autofluorescence over time
•
Calculate absolute ion concentrations (pixel by pixel) via titration calibration (known ion concentrations
applied to the cells – in situ – or in solutions – in vitro or equation calibration where possible [Fura-2,
Indo-, SNARF]
•
Calculation of R eliminates artifacts and uncertainties caused by
−
inhomogenous dye distribution
−
photobleaching
−
may be applied with moving cells
4.6.12.5
Ratiometric Dyes - Online Ratio
R(t1) = F1(t1) / F2(t1), R(t2) = F1(t2) / F2(t2) ...
Fig. 4-139
Ratiometric Dyes – Online ratio
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......