LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Image Optimization
Carl Zeiss
03/06
B 45-0021 e
4-351
If the image is not bright enough, it appears blue
on the screen.
•
On the
Channel Settings
panel of the
Scan
Control
window, set the PMT (photomultiplier)
gain with the
Detector Gain
slider.
−
The image should not have more than a
trace of red.
−
This adjustment is very sensitive. Try using
the left and right arrows to make the
adjustment instead of dragging the slider
bar.
•
To adjust the black level (background), use the
Ampl. Offset
slider so that areas without
picture content just show a trace of blue.
•
If necessary, re-amplify brightness with the
Ampl. Gain
slider.
Do not change the
Ampl. Gain
setting
unless the settings made so far are
insufficient for image optimization.
•
In the Color Palette List panel of the
Color
Palette
window, click on
No Palette
.
−
This deselects the
Range Indicator
and
activates the new presentation.
•
In the
Scan Control
window, click on the
Stop
button.
−
This stops the continuous scan.
If you use the
Range Indicator
for image
optimization, it may happen that the
ranges marked in the
Range Indicator
will vary when the channel color is
changed.
Fig. 4-331
Image Display window
Fig. 4-332
Image Display window
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......