IMPORTANT NOTES FOR CHAPTER 4
LSM 510 META
Carl Zeiss
03/06
B 45-0021 e
4-I
IMPORTANT NOTES FOR CHAPTER 4
Bleach series
When doing bleach series on LSM 510
DuoScan
systems, the time stamps and
bleach markers indicated in the images will depend on the start position of the
scanners in the image frame. Since these are not identical for the imaging and the
photomanipulation scanners, sometimes imaging seems to start too early after a
bleach event. Ignore the last two digits in the time stamps and bleach markers in
this case.
Macro
When using the modify images macro to convert time series data into Z stacks, an
arbitrary section step of 1 µm is applied to the stack. This will also lead to an
arbitrary time stamp when reconverting the Z stack into a time series. The order of
the images however is preserved under all conditions.
Piezo objective
focus units
The piezo objective focus units 1325-210, 1325-212 and 1325-213 with an
increased travel range of 250 µm are controlled by a 14 bit logic. Due to the
increased travel range, this implies a smallest step of 15 nm, in contrast to former
fine focus solutions with a smaller travel range but a smallest step size of <10 nm.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......