OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Process Menu
LSM 510 META
DuoScan
4-152
B 45-0021 e
03/06
4.6.12.6
Ratiometric Dyes - Calibration
•
Subtract background/autofluorescence images from raw images to obtain
R
korr
[=F
1
-F
1Background
)/(F
2
-F
2Background
)]
when calibration reference is not obtained with the experimental sample (in situ)
•
Calculate ratio R
•
Perform equation- or titration calibration (compare R with a calibration curve -> titration calibration or
put R values in calibration formula)
Fig. 4-140
Ratiometric Dyes - Calibration