OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-75
Speed
panel
•
Select the
Scan Speed
from the 13 preset steps
via slider or input box. Recommended: 7 for the
first scan. A click on the
Max
button sets the
maximum speed for the current zoom.
−
The Scan Speed determines the Pixel Time. In the case of different image formats, the Pixel Time is
constant for the same Scan Speed, but the Scan Time is different.
−
Pixel Time dwell time of the laser beam on the pixel
−
Scan Time duration of the acquisition for the entire frame
Fast XY only for fast image acquisition during parameter setup.
Pixel time and scan time will be shown. Fast XY = speeds 8 – 13 (depending on zoom),
average = 1, max. resolution: 512 x 512 pixels.
Speed:
1 ... 8
9
10
11
12
13
Zoom:
0.7 1.0 1.4 2.5 3.6 5.7
Pixel Depth, Scan Direction & Scan Average
panel
•
Select
8 Bit
or
12 Bit
Data Depth, i.e. 256 or
4096 gray values.
•
Select the
Unidirectional
or
Bi-directional
Scan Direction.
−
Unidirectional: The laser scans in one
direction only, then moves back with beam
blanked and scans the next line.
−
Bi-directional: The laser also scans when
moving backwards, i.e. the Scan Time is
halved.
−
The pixel shift between forward and backward movement (double image) resulting from bi-
directional scanning must be corrected via the
Scan Corr X
and
Y
sliders. Zero° rotation requires
correction in the X-direction, 90° rotation must be corrected in the Y-direction. If the image was
rotated, correction is required in both coordinates. Correction is performed on-line in the
continuous scan mode (
Cont.
button). The size of the shift depends on the Scan Speed. For
automatic scan correction, click on the
Auto
button.
•
Select the
Line
or
Frame
mode for averaging.
•
Select the desired scan average method
Mean
or
Sum
in the
Method
selection box.
•
Select the desired scan average from the available values
2
,
4
,
8
and
16
in the
Number
selection box
or
Continues
(only for
Frame
average mode).
The greater the number of averages selected for
Mean
average
Method
, the better the image
quality will be; the scanning time will be prolonged accordingly.
Fig. 4-53
Speed panel
Fig. 4-54
Pixel Depth, Scan Direction &
Scan Average panel
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......