LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-67
Control buttons
Close
Closes the
Configuration Control
window.
Spectra
The
Spectra
button opens the
Detection Spectra & Laser Lines
window (see
page 4-53).
Laserline
The
Laserline
button opens the
Wavelength Switch Control
window (see
page 4-54).
Config
The
Config
button permits existing track configurations to be loaded, stored
under any name, or deleted (see page 4-54).
(2)
Beam path - Excitation
•
On the
Beam Path and Channel Assignment
panel, click on the
Excitation
button.
−
This opens a dialog box of all available lasers with their wavelengths and their usable Acousto-
Optical Tunable Filters (AOTF) attenuation.
•
To select the desired laser line, activate the check box for
Line Active
.
•
Use the
Transmission [%]
slider to set the utilizable laser intensity (recommendation: start at 50 %).
−
The transmittance of the Acousto-Optical Tunable Filter (AOTF) changes accordingly.
•
This allows you to adapt the laser intensity very sensitively to the job. Activate the check box for
Line
Active
.
−
This activates the selected laser power for use. This is indicated by the
Laser Power
displaying
lamps (status display green / gray).
By clicking on the
Excitation
button you can check at any time which lasers are available for
active operation.
If you deactivate
Line Active
, the laser wavelengths are deselected by means of the Acousto-
Optical Tunable Filters (AOTF), i.e. these lasers change into standby status.
If you interrupt your work with the LSM for a break, it is recommended not to switch the
Enterprise and argon lasers off by hardware action, but to put them into standby status as
described.
Excitation filters, emission filters, HFT main dichroic beam splitters and NFT secondary dichroic
beam splitters can be switched online, channels (PMT photomultipliers) only off-line.
(3)
Beam path - HFT main (dichroic) beam splitter
•
On the
Beam Path and Channel Assignment
panel, click on the HFT main dichroic beam splitters
.
−
This opens a graphical pop-up window of all beam splitters available.
•
To select a beam splitter, click on the respective line of the list.
−
The selected beam splitter moves into the beam path.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......