LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Display and Analysis of Images
Carl Zeiss
03/06
B 45-0021 e
4-277
Histogram functions
Skip Black
button
Ignore black pixels (gray value 0) in the histogram.
Skip White
button
Ignore white pixels (gray value 255 or 4096) in the histogram.
Statistics
Displays statistical parameters (Mean Intensity, Standard Deviation, Number
of Pixels and Size of Area) in a additional table. Area measurements of very
small areas (< 10 pixels) give only approximate values
Step
input box
Set the number of intensity steps which shall be displayed in the diagram.
Step 1 corresponds to 256 intensity steps, Step 64 to 4 intensity steps in case
of 8 bit images. Reduction is made by averaging.
Show Image
button
Shows the image in the
Image Display
window beside the histogram.
Show Table
button
The histogram is shown as a table at the bottom left of the
Image Display
window.
Copy Table
button
The histogram table is copied to the clipboard.
Save Table
button
The histogram table can be stored as a text file (extension *.txt).
Area functions
Area
button
Interactive definition of area for size and intensity measurement.
Save Values
button
Copies area values to the clipboard (only available if the
Area
button is
activated).
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......