OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Process Menu
LSM 510 META
DuoScan
4-156
B 45-0021 e
03/06
−
The amplitude of the fitted curve (which equals the mobile fraction)
I1 mobile fraction
.
−
The fitted parameter
T1
(seconds).
−
The rate constant for the exchange of molecules between the bleached region and the surrounding
area
K
(per second).
−
The part of the immobile fraction of the protein
I delta immobile fraction.
Show Table
opens a further table beneath the results. It shows all intensity values being corrected for
background intensity and intensity loss of the reference region.
The values can be saved as a text file (
Save Table
) or copied into Excel via clipboard (
Copy Table
).
The data can be normalized optionally when marking the checkbox
Normalize
in the Kinetic Analysis
Diagram toolbar.
The calculation of the parameters is based on the same ROIs unless other ROIs or moved ROIs
are selected again. The Kinetic Display is always available once the analysis has been performed.
The analysis is not stored with the image.
Note that this modeling is a very basic approach to your experiment. It offers a first hint on the
possible presence of only one or, in case of a bad fit, more than one mobile fractions of the
labeled protein within the cell or cell compartment examined.
The half time of the recovery can be calculated using the following formula:
1
T
5
.
0
ln
t
half
−
=
Fig. 4-143
Image window displaying the analysis of FRAP data using a mono
exponential fit
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......