Routine Mode
LSM 510
DuoScan
Carl Zeiss
Application of Examination Methods in …
LSM 510 META
DuoScan
6-16
B 45-0021 e
03/06
(c)
Starting a time series via the time
•
To start a time series via the time, the
Start
button must be pressed first.
−
Waiting for Start Time
will be displayed in the status line. The time series is started when the
starting time has been reached.
−
The time series is interrupted when the end time has been reached.
If the entered
Number
of cycles has been processed, the time series is finished.
If the number of cycles has not yet been processed, the time series is only interrupted.
Waiting for Start
Time
is displayed in the status line. The time series can now be continued by entering a new start time or
finished via
Stop
.
The end time for the time series can be changed online.
(d)
Input of the number of scan cycles
The currently set number of scan cycles is displayed in the
Number
input box of the
Start & End
panel.
This number can be changed, if required.
•
Set the required
Number
of scan cycles using the sliders or the arrow buttons.
(2)
Cycle Delay / Time Interval panels
Depending on the Examination Method exported
from the Expert Mode, the time series interval is
defined either as a
Cycle Delay
or
Time Interval
.
Accordingly, either the
Time Delay
panel or the
Time Interval
panel is displayed in the
Routine
Mode - Time Settings
window.
Time Delay
is the interval between the end of one
scan process and the beginning of the next.
Time Interval
is the interval between the
beginning of one scan process and the beginning
of the next.
The
Scan Interval
panel permits the intervals to
be activated and changed.
The following functions are available:
Interval
buttons
Activation of the interval set for the time series for the relevant button.
Time
input box /
Determination of the interval for the currently selected time button.
arrow buttons / slider
Unit
buttons
Selection of the time unit:
min
,
sec
or
ms
.
Trigger in
field
Display of the trigger keys (1-4) for the in signal.
Trigger out
field
Display of the trigger keys (1-4) for the out signal.
Fig. 6-21
Time Delay panel
Fig. 6-22
Time Interval panel
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......