LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-89
Auto Z Corr.
The function
Auto Z Correction
allows a linear variation of Detector Gain, Ampl. Offset, and Ampl. Gain
values between the different slices of a stack.
•
Click on the
Auto Z
button, the
Auto Z
Brightness Correction
panel opens.
The buttons
Set A
and
Set B
permit definition of
two distinct gain / offset / AOTF settings at two
different Z positions A and B.
Pressing the
Move A
and Move
B
buttons permits
the defined Z-position to be directly approached.
The
Enable test
check box permits simulation of
the value changes for
Detector Gain
,
Ampl.
Offset
,
Ampl. Gain
and
Attenuation
in the
Scan
Control
window without the scanners being in
operation.
If a Z Stack is performed and the
Auto Z Brightness Correction
window is opened this correc-
tion is automatically performed equal whether the
Enable test
box is enabled or disabled.
•
Use the focusing drive to set the Z-position where the brightness level correction is to be started.
•
In the
Scan Control
window, set the initial values for
Detector Gain
,
Ampl. Offset
and
Ampl. Gain.
If required, start the continuous scan procedure for this purpose. Click on the
Set A
button.
•
Use the focusing drive to set the Z-position where the brightness level correction is to be ended.
•
Set the end value for
Detector Gain
,
Ampl. Offset
and
Ampl. Gain
in the
Scan Control
window
.
Click on the
Set B
button.
•
If required, check the change of the set values by activating
Enable test.
After the start of the scan procedure, the brightness level values are linearly interpolated between the
defined references A and B.
Note that the total Z range where the interpolation takes place can exceed the Z reference A
and B.
Fig. 4-69
Z Brightness level control panel
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......