Installation
•
Upgrade a DigiVAC Standard to a
Watkiss DigiVAC Service Manual - Issue 2 - 29/11/04
119
6. Screw two stand-off posts into the holes proved in the base module (see
Figure 7:5 Stand-off Posts (circled)
7. Take the DigiVAC CPU 20-way cable and plug it into the DigiVAC Base
PCB. Thread the cable behind the PCB and downwards so that it comes
out at the bottom of the PCB. (You may need to ease the PCB off its stand-
off posts in order to do this).
8. Fix the CPU box in place by inserting two screws (with washers) into the
stand-off posts fitted to the base module (see Figure 7:5). The 20-way CPU
cable fits behind the CPU box.
Figure 7:6 CPU Box and Connections
20-Way CPU Cable
CPU Box
20-Way CPU Cable
50-Way Touch Screen Cable