Without the column, backpressure is usually less than 0.34 MPa (50 psi),
depending on the flow rate and tubing ID. Note that the semi-micro flow
cell will develop approximately 1.03 MPa (150 psi) backpressure at
1.0 mL/min.
If the backpressure remains high, remove the fittings and tubing from part
of the flow path, section by section; starting at the waste line, work back
“upstream” until you identify the section causing the substantial
backpressure. Replace tubing and fittings as necessary.
•
Periodically check the flow cell tubing connections for leaks or restrictions.
Replace tubing and fittings as necessary.
•
Periodically repassivate stainless steel systems. This is especially important
when operating with extreme pH levels (which will shorten the life of the
passivation finish) or after a major disassembly. For instructions, see
Passivation of Stainless Steel Systems and Components
(Document No.
031152).
•
Monitor baseline noise. If the noise increases above an acceptable level,
especially after a lamp(s) has been in service for an extended length of time,
replace the lamp.
To check the lamp lifetime:
1. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
2. Select
UV
, and then select
UVLampAge
and
VisLampAge
. If a lamp has
been in service for more than 2000 hours, replace it. See
or
for instructions on replacing the deuterium lamp or tungsten
lamp, respectively.
•
Check the deuterium lamp intensity approximately every 6 months:
1. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
2. Select
UV
, and then select
LampIntensity
. The reading should be above
10 million counts for a standard or semi-preparative cell or above 3
million counts for a semi-micro cell.
3. If the lamp intensity reading is too low, calibrate the intensity: In the
Command window or Commands dialog box, select
UV_Calibration
,
select
Intensity,
and execute the command.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...