![Thermo Scientific 074114 Скачать руководство пользователя страница 126](http://html1.mh-extra.com/html/thermo-scientific/074114/074114_operators-manual_1105368126.webp)
Dionex PDA Operator’s Manual
118
Doc. 065378-01 1/12
The table below describes the pin assignments for the relay and TTL
connector. Relays 1 and 2 can be configured in Chromeleon to switch any
low-voltage control. The switched current must be less than 200 mA and
42 V peak.
B.7 Connecting to the Chromeleon PC
B.7.1
Before You Begin
Verify that the following tasks have been completed:
•
Chromeleon 7 or Chromeleon 6.8 is installed on the computer.
•
The Chromeleon 7 or Chromeleon 6.8 license is installed.
NOTE Thermo Fisher Scientific strongly recommends
installing Chromeleon
before
connecting the Dionex
PDA to the computer. When Chromeleon is installed
first,
USB
driver
information
is
loaded
automatically.
Pin
Assignment
Input Pin Function
1
Relay 1 output
Active
2
Relay 1 output
Ground
3
Relay 2 output
Active
4
Relay 2 output
Ground
5
TTL 1 output
Active
6
TTL 2 output
Active
7
TTL 1 input
Active
Autozero
8
TTL 2 input
Active
Reserved
9
TTL 3 input
Active
UV lamp on/off
10
TTL 4 input
Active
Visible lamp on/off
11
TTL input or output
Ground 1
12
TTL input or output
Ground 2
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...