Dionex PDA Operator’s Manual
82
Doc. 065378-01 1/12
•
Flow cell contains air bubble or old mobile phase
Flush the cell with the mobile phase for the application.
•
Cell not clear
Clean the cell (see
If the detector fails the wavelength calibration again, replace the flow cell
windows (see
). Contaminated flow cell windows can significantly
increase baseline noise and cause wavelength calibration to fail.
4.9
Wavelength Verification Fails
•
Detector wavelength requires recalibration
Open the Dionex PDA Wellness panel and run the wavelength calibration
procedure (see
•
No solvent flow through cell
Check the Chromeleon Audit Trail. If the “Not enough light to perform
operation.” message is displayed, flush the cell with deionized water or
methanol until the autozero routine can be performed without error.
•
Lamp(s) needs replacement
Check the number of hours the lamps have been in service:
1. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
2. Select
UV
, and then select
UVLampAge
and
VisLampAge
. If a lamp has
been in service for more than 2000 hours, replace it. See
or
for instructions on replacing the deuterium lamp or tungsten
lamp, respectively.
•
Solvent absorbs at the wavelength reported for verification
Pump deionized water or methanol through the system during the wavelength
verification test.
•
Filter paddle position error
Check the Chromeleon Audit Trail. If the “Filter wheel sensor could not
detect blocked position.” or “Filter wheel sensor could not detect open
position.” message is displayed, see
for instructions.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...