3 • Operation and Maintenance
Doc. 065378-01 1/12
47
Reference Wavelength
The reference wavelength is the center wavelength of the reference intensity
determination. Absorbance is measured by comparing the reference intensity to
the sample intensity. Selecting a reference wavelength implies a reference mode
for each of the single wavelengths being collected. There are two types of
reference mode: “active” or “off.”
•
To
select
the active reference mode, enter a wavelength within the selected
sample scanning range of the Dionex PDA. The selected wavelength will be
used as the reference wavelength. The reference intensity will vary over time,
as the signal intensity changes.
For maximum compensation, select a reference wavelength from the same
lamp as the sample wavelength. For example:
If the sample wavelength is 272 nm, you might select a reference
wavelength of 325 nm because both wavelengths are within the UV
spectrum (190 to 380 nm) of the deuterium lamp.
Figure 3-6. Effect of Bandwidth on Baseline Noise
-0.0500
-0.0400
-0.0300
-0.0200
-0.0100
0.0000
0.0100
0.0200
0.0300
0.0400
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
4.00
4.50
5.00
mAU
min
1
2
3
3
2
1
WVL:254 nm
Bandwidth = 1
Noise = 26
P
AU
Bandwidth = 4
Noise = 16 AU
P
Bandwidth = 8
Noise = 14
P
AU
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...