4 • Troubleshooting
Doc. 065378-01 1/12
65
To troubleshoot:
During data acquisition, no other commands can be executed. Wait until data
acquisition is completed before attempting to issue any other command.
To troubleshoot:
Wait until the autozero routine is completed before attempting to issue any
other command.
To troubleshoot:
This message appears if the Dionex PDA fails to respond when the computer
attempts to communicate with the detector. Shut down the Chromeleon
instrument controller or server and turn off the Dionex PDA power; wait
30 seconds, and then power up both the instrument controller or server and
the detector. If the error message appears again, contact Thermo Fisher
Scientific for assistance.
Can’t execute calibration command during data acquisition.
Can’t execute diagnostic command during data acquisition.
Can’t execute x command during data acquisition.
Command not executed because an Autozero command is
already running.
Data is unavailable.