Dionex PDA Operator’s Manual
100
Doc. 065378-01 1/12
15. Replace the lamp cover.
16. Thread the flow cell inlet tubing through the tubing guide on the lamp cover
and reinstall the tubing bracket.
17. Replace the flow cell cover and reconnect the inlet and outlet tubing.
18. Replace the detector front cover.
19. Turn on the detector power.
20. Start the Chromeleon instrument controller or server.
21. Reset the lamp lifetime:
a. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
b. Select
UV
, and then select the
UVLampAge
command.
c. Reset the lamp age to zero.
d. Execute the command.
NOTE It is normal for the baseline to drift for several hours
after a new deuterium lamp is installed.
22. Run the wavelength calibration and wavelength verification test for the new
Do not pinch the UV lamp connector wires under the lamp cover.
Do not pinch the tubing under the flow cell cover.