4 • Troubleshooting
Doc. 065378-01 1/12
73
4.2
ALARM LED Is Lighted
The leak sensor in the drip tray may have been triggered. Check the Chromeleon
Audit Trail for a leak-related error message. Find and eliminate the source of the
leak.
After fixing the leak and drying the sensor, clear the
ALARM
LED.
•
In the Chromeleon 7 Command window (see
): Click
UV
, and then
execute the
ClearAlarmLED
command.
•
In Chromeleon 6.8, click the
Clear Alarm LED
button on the Control panel.
–or–
In the Commands dialog box (see
), click
UV
and then execute the
ClearAlarmLED
command.
4.3
Lamp Does Not Light/Lamp LED Is Flashing Rapidly
•
Lamp default setting is not in effect
When the factory default settings are in effect, the deuterium and tungsten
lamps turn on automatically each time the detector power is turned on. If the
required lamp does not turn on automatically, you can turn it on from
Chromeleon or the TTL input.
To restore the default setting for a lamp:
1. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
2. Select
UV
, and then execute the required command:
•
UVLampPowerOnSetting
(for the deuterium lamp)
•
VisLampPowerOnSetting
(for the tungsten lamp)
•
Lamp is old or burned out
Replace the lamp. See
or
for instructions on replacing
the deuterium lamp or tungsten lamp, respectively.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...